PCT17144智能微量残炭测定仪 一、概述 本仪器是按照GB/T 17144《石油产品残炭测定法》、《ASTM D4530》和《ISO 10370:1993》的标准要求设计制造的。具有外形美观,操作方便等特点,其测试结果与GB/T268(康氏法)测试结果*。 PCT17144智能微量残炭测定仪 二、主要技术参数 1、输入电源:AC 220V±5% 50Hz 2、加热功率:950W 3、控温点:500℃ 4、控温精度:±2℃ 5、zui多可做试样数:12个 6、工作压力:0.15Mpa 7、气源压力:一级压力zui小2MPa ,二级压力zui大0.5Mpa 8、氮气流量:150mL/min 600mL/min (按照GB/T 17144-1997) 9、工作时间:约120min 10、工作环境温度:≥25℃ 11、工作环境相对湿度:<85%RH 12、外形尺寸:380×410×570mm(长×宽×高) PCT17144智能微量残炭测定仪 三、主要特点 仪器外形及结构如上图所示。与传统的康拉特逊残炭测定仪相比本仪器具有以下几项特点: 1、只需少量的试样即可进行试验: 残炭量(%) <1 >1-5 >5 试样量(g) 5.0±0.5 0.5±0.1 0.15±0.05 2、可同时测定12个样品。 3、采用氮气保护气流,确保试验条件不变能够获得良好的重复性和再现性。 4、采用气管导出法排放试验尾气,同康氏法测定仪相比对实验场所的要求更加自由灵活。 5、采用轴流风机进行冷却,有助于降低仪器温度。 6、操作仅需按动“启动”键,一切均由仪器自动进行测试,正常时无需任何操作和判断,试验结束后,蜂鸣器提示试验结束。 7、显示直观明了,方便观察。 PCT17144 inligent micro carbon residue tester One, overview The apparatus is in accordance with the GB/T 17144 "petroleum products - Determination of carbon residue," "ASTM D4530" and "ISO 10370:1993" standard requirements of design and manufacturing. Has the advantages of beautiful appearance, convenient operation, the test results and the GB/T268 (Kang's method) consistent test results. Two, the main technical parameters 1, the input power supply: AC 220V + 5% 50Hz 2, the heating power: 950W 3, temperature 500 DEG C: 4, temperature control precision: - 2 DEG C 5, most can be used as specimen number: 12 6, working pressure: 0.15Mpa 7, air pressure: the pressure minimum level 2MPa, two level of maximum pressure for 0.5Mpa 8, the nitrogen flow rate: 150mL/min 600mL/min (according to GB/T 17144-1997) 9, working time: about 120min 10, the working environment temperature: 25 C 11, the environmental relative humidity: <85%RH 12, size: 380 * 410 * 570mm (length * width * height) Three, the main features The shape and structure of the instrument are shown. And Kanglatexun carbon residue tester compared with the traditional instrument has the following features: 1, only a small amount of sample can be tested: The amount of residual carbon (%) < 1 > 1-5 > 5 Sample size (g) 5 + 0.5 0.5 + 0.1 0.15 + 0.05 2, the simultaneous determination of 12 samples. 3, protected with nitrogen gas flow, ensure the constant test conditions to obtain good repeatability and reproducibility. 4, use tracheal derived method of emission test exhaust, with Kang's method determination instrument compared to the experimental location requirements more free and flexible. 5, adopt a shaft for cooling flow fan, helps to reduce the temperature of the instrument. In 6, operation only needs to press the "start" button, all are the automatic test instrument, the normal without any operation and judgment, after the test, the buzzer tip end test. 7, display intuitive, convenient for observation. |